Thermosel System | Brookfield
The difficulty with viscosity measurements of hot melts and liquids at elevated temperatures has been in maintaining accurate temperature control that is consistent from sample to sample so that meaningful data could be obtained. The Brookfield Thermosel solves this problem by providing a stable, precisely controlled sample environment. This, together with the inherent accuracy of the Brookfield Viscometers, is fundamental to the Thermosel System, which produces viscosity measurements that are not only accurate but entirely reproducible.
Several factors contribute to the stable environment:
- Non-fluctuating temperature control.
- Small sample volume and insulated sample chamber which reduces temperature gradients within the sample.
- The rotating spindle, which acts as a built-in stirring device.
- The test procedure is quite straightforward. Once familiar with the system, unskilled operators can easily produce accurate, reproducible data.
FEATURES & BENEFITS
- Provides control of sample temperature up to +300°C
- Programmable Controller offers single set point or up to 10 programmable set points. Temperature ramping between set points is possible if used with RheocalcT Software
Note: Requires optional cable HT-106 when QC control with RheocalcT is implemented. - Compatible with standard Brookfield Viscometers and DV3T Rheometers
Note: requires optional cable DVP-141 when controlled by DV3T Rheometer - Supplied with reusable stainless steel sample container and five disposable aluminum sample chambers
- Outputs: 0-4 Vdc for chart recorder; RS-232 for use with DV3T and/or RheocalcT
DISPOSABLE SPINDLE/CHAMBERS OPTION
Disposable sample chamber with Disposable Spindle SC4-27D* is ideal for asphalts or any difficult-to-clean material. Requires special adapter (SC4-DSY).
*Optional SC4-27D is available in quantities of 100, (SC4-27D-100).
EZ-LOCK OPTION**
The Thermosel is also available to purchase with an EZ-Lock spindle couple system. This option allows spindles to be quickly attached and removed, and may also help prevent against damage that can occur from frequent spindle changes or multiple users. Existing Thermosels can also be retrofitted. Please contact Brookfield or an authorized dealer for details and part numbers.
**Note: The EZ-Lock option must be purchased for use with all “EXTRA” series Viscometers/Rheometers as well as all other instruments that have an EZ-Lock spindle coupling system.
What’s Included
Alignment Bracket
Thermo Container with safety guard and insulating cap
One Removable Sample Chamber
Five Disposable Sample Chambers
Extracting Tools
Temperature Controller with an RTD probe
SPECIFICATIONS
Thermosel Adapter Viscosity Ranges cP (mPa·s) | |||||
SPINDLE SAMPLE VOLUME SHEAR RATE (SEC-1) † MODEL |
SC4-18 8mL 1.32N | SC4-31 10mL .34N | SC4-34 9.5mL .28N | SC4-21 8mL .93N | SC4-27* 10.5mL .34N |
LVDV3T | 1.2-30K | 12-300K | 24-600K | Not applicable for historical | |
LVDV2T | 1.5-30K | 15-300K | 30-600K | reasons. However, it is possible | |
DV1MLV | 3-10K | 30-100K | 60-200K | to use the above spindles | |
LVDVE | 3-10K | 30-100K | 60-200K | with any of these instruments. | |
LVT | 5-10K | 50-100K | 100-200K | Digital Viscometers/Rheometers | |
RVDV3T | 20-500K | 100-2.5M | |||
RVDV2T | 25-500K | 125-2.5M | |||
DV1MRV | 50-170K | 250-830K | |||
RVDVE | 50-170K | 250-830K | |||
RVT | will automatically calculate | 50-100K | 250-500K | ||
HADV3T | viscosity. Please contact | 40-1M | 200-5M | ||
HADV2T | Brookfield or an authorized | 50-1M | 250-5M | ||
DV1MHA | dealer if you require | 100-300K | 500-1.7M | ||
HADVE | Information on viscosity range. | 100-300K | 500-1.7M | ||
HAT | 100-200K | 500-1M | |||
HBDV3T | 160-M | 800-20M | |||
HBDV2T | 200-4M | 1K-20M | |||
DV1MHB | 400-1.3M | 2K-6.7M | |||
HBDVE | 400-1.3M | 2K-6.7M | |||
HBT | 400-800K | 2K-4M |
Thermosel Adapter Viscosity Ranges cP (mPa·s) | |||
SPINDLE SAMPLE VOLUME SHEAR RATE (SEC-1) † MODEL |
SC4-28 11.5mL .28N |
SC4-29 13mL .25N |
HT-DIN-81** 7mL 1.29N |
LVDV3T | 1.0-10K | ||
LVDV2T | Not applicable for historical reasons. | 3.4-10K | |
DV1MLV | See note above. | 3.4-10K | |
LVDVE | N/A | ||
LVT | 5.7-10K | ||
RVDV3T | 200-5M | 400-10M | 14.6-10K |
RVDV2T | 250-5M | 500-10M | 36.5-10K |
DV1MRV | 500-1.7M | 1K-3.3M | 36.5-10K |
RVDVE | 500-1.7M | 1K-3.3M | N/A |
RVT | 500-1M | 1K-2M | 36.5-10K |
HADV3T | 400-10M | 800-20M | 29.2-10K |
HADV2T | 500-10M | 1K-20M | 73.0-10K |
DV1MHA | 1K-3.3M | 2K-6.7M | 73.0-10K |
HADVE | 1K-3.3M | 2K-6.7M | N/A |
HAT | 1K-2M | 2K-4M | 73.0-10K |
HBDV3T | 1.6K-40M | 3.2K-80M | 116.8-10K |
HBDV2T | 2K-40M | 4K-80M | 292.0-10K |
DV1MHB | 4K-13.3M | 8K-26.7M | 292.0-10K |
HBDVE | 4K-13.3M | 8K-26.7M | N/A |
HBT | 4K-8M | 8K-16M | 292.0-10K |
M=1 million K=1 thousand N=RPM cP=centipoise mPa·s=Millipascal·seconds
† Spindle SC4-18 1.32×10(rpm)=13.2 sec-1
**The 81 spindle, Part No. HT-DIN-81, works in an HT-2 or HT-2DB chamber.